MJ-560B2 Anesthesia Machine, designed with state-of-the-art technologies, is one of high utility with excellent performance and complete functions. Both outstanding and inexpensive, it has the best cost performance that can satisfy any customer.
The Ventilator adopts microcomputer control with high-definition 5.7" LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.
The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose onekindof gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
Reliable pneumatic transmission system.
Economic and proper configuration of technologies, optional as required by the Customer.
Modularized design and upgrading. Upgrading is available as required by the Customer.
Specification
Main Body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
Scope of application | Adult, child |
Gas Source | O2, N2O:0.28-0.6MPa N2O: 0.28~0.6MPa |
Flow Meter | O2: 0.05-10L/minN2O: 0.05~10L/min |
O2, N2Olinkage and N2O Stopper | When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off. |
Flow Rate of Rapid Oxygen Supply | 25-75L/min |
Low Oxygen Pressure Alarm | There will be sound alarm when the oxygen pressure< 0.2 MPa |
Vaporizer | It has the function of automatic compensation based on pressure, temperature and flow rate. The regulation range of evaporator concentration is 0-5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.And vaporizers that are imported with original packaging are also available. |
Low oxygen pressure alarm | There will be sound alarm when the oxygen pressure<0.2Mpa |
Respiratory Circuit | Working Mode: all-close, semi-close, semi-open APL: ≤12.5kPa |
Respiratory Bellows | Bellows for adults, bellows for children Tidal volume range: 0-1500ml |
Display Mode | High-definition 5.7'' LCD screen display |
Ventilation Mode | IPPV, SIPPV, SIMV, MANUAL |
Ventilation Function | PEEP, SIGH, IRV |
Ventilation Parameters | Tidal Volume: 50-1500ml Rate: 2-99bpm SIMV: 2-20bpm I:E 2:1-1:8 Inspiratory trigger pressure -10-20cm H2O PEEP 2~20cmH2O Pressure limitation 10~60cmH2O SIGH 1.5 times the inspiratory time/100bpm |
Parameters for Ventilation Monitoring | Tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure -time waveform, flow -time waveform, PEEP, inspiratory trigger pressure |
Airway Pressure Alarm | Upperlimit setting range 5~60cmH2O Low limit setting range 0~50 cmH2O |
Minute Ventilation Volume Alarm | Upperlimit setting range 1~20 L/min Low limit setting range 0~19L/min |
Tidal volume | Upperlimit setting range 100~1500ml Low limit setting range 0~1400ml |
Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 cmH2O (Suffocation alarm, Power alarm, Intubation off alarm, battery voltage alarm) |
Power | AC220V50Hz(UPS, Storage battery) |
*(Mechanical Arms) | |
*(End-tidal CO2 monitoring) | |
*(Specific interface of waste anesthesia gas) |